Skip to main content

Range modulation in proton therapy planning: a simple method for mitigating effects of increased relative biological effectiveness at the end-of-range of clinical proton beams

Abstract

Background

The increase in relative biological effectiveness (RBE) of proton beams at the distal edge of the spread out Bragg peak (SOBP) is a well-known phenomenon that is difficult to quantify accurately in vivo. For purposes of treatment planning, disallowing the distal SOBP to fall within vulnerable tissues hampers sparing to the extent possible with proton beam therapy (PBT). We propose the distal RBE uncertainty may be straightforwardly mitigated with a technique we call “range modulation”. With range modulation, the distal falloff is smeared, reducing both the dose and average RBE over the terminal few millimeters of the SOBP.

Methods

One patient plan was selected to serve as an example for direct comparison of image-guided radiotherapy plans using non-range modulation PBT (NRMPBT), and range-modulation PBT (RMPBT). An additional plan using RMPBT was created to represent a re-treatment scenario (RMPBTrt) using a vertex beam. Planning statistics regarding dose, volume of the planning targets, and color images of the plans are shown.

Results

The three plans generated for this patient reveal that in all cases dosimetric and device manufacturing advantages are able to be achieved using RMPBT. Organ at risk (OAR) doses to critical structures such as the cochleae, optic apparatus, hypothalamus, and temporal lobes can be selectively spared using this method. Concerns about the location of the RBE that did significantly impact beam selection and treatment planning no longer have the same impact on the process, allowing these structures to be spared dose and subsequent associated issues.

Conclusions

This present study has illustrated that RMPBT can improve OAR sparing while giving equivalent coverage to target volumes relative to traditional PBT methods while avoiding the increased RBE at the end of the beam. It has proven easy to design and implement and robust in our planning process. The method underscores the need to optimize treatment plans in PBT for both traditional energy dose in gray (Gy) and biologic dose (RBE).

Background

Proton beam therapy (PBT) has emerged as an important advance in radiation therapy, particularly for children and young adults. Although there are data supporting equivalent rates of cure using equivalent doses of proton versus photon treatments, the decreased dose to nearby OARs and substantial reduction of irradiated tissue volume is a promising strategy to address acute and late injury to normal tissues from therapy [1]. Additionally, a matched pair analysis [2] supports a range from no difference to fewer second malignancies in patients treated with protons versus photon therapy. Similar data comparing photon CSI to proton CSI from multiple institutions also suggests a clear correlation of irradiated volume to risk of second malignancy [36].

Increased tissue effects at the end of the spread out Bragg Peak (SOBP) is a defined, measured biologic phenomenon [710]. The potential for unintentional tissue injury due to putative increased relative biological effectiveness (RBE) at the distal edge of a proton beam is an issue that must be addressed during the planning process [8, 1113]. The existing models suggest there may be a 5-10% increase in biological effect at the most distal portion of the SOBP relative to the plateau and an extension of effective proton range by 1–2 mm independent of fractionation and tissue type, while noting “There are no proton RBE values based on human-tissue response data, despite clinical experience of the treatment of more than 50,000 patients”. [ICRU-78 [14]; Section 2.4] Current areas of research are exploring how the RBE of protons also varies with fraction size and how it affects range [15, 16].

In recognition of this dilemma, some current pediatric treatment protocols mandate multi-beam proton plans with the rationale that multiple beam entry directions will reduce the dose to critical structures in close proximity to the tumor. However, adding more beams may be detrimental in some cases or may simply increase logistics without therapeutic benefit. An example is the re-treatment of a brain tumor in which the brainstem, skin, and surrounding tissues have already acquired a significant dose from prior treatment. There may not be an optimal beam entry direction, let alone multiple entry directions in such cases. To address the issue of a potential increased RBE at the distal edge of a proton beam, another approach chooses to stop the beam beyond, rather than in, the OAR in order to place the distal portion of the SOBP in less vulnerable tissue downstream. This requires acceptance of the OAR receiving full dose uniformly as a safety measure in order to avoid a potentially serious, but poorly quantifiable, complication. Using mixtures of these methods, it is possible that OARs would receive more dose than they would from treatment with an alternative modality, such as IMRT. Both of these strategies keep practitioners from using proton technology optimally. Range modulation has been deployed across a number of different tumor types for various planning scenarios since August 2010 when first developed. In order to make this comparison valid, one patient was selected and three plans were constructed to demonstrate the use of this new technique.

Methods

This manuscript describes a novel technique to mitigate issues related to increased RBE at the distal edge of the SOBP by spoiling the distal falloff with existing patient specific device (PSD) sets and beam angles. We call this technique “range modulation” or “range mod”. This is accomplished by splitting the dose planned for a beam in half, shown in Figure 1, and then delivering half the dose as planned and other half of the dose with an identical beam whose range has been modified by 3 mm [see reference on treatment system ref [13], 3 mm is half the 6 mm spacing between pristine peaks in the SOBP for our beam delivery system and comparable to the potential 1–2 mm increase in range due to RBE). If a single beam is being used for a plan to a significant dose, three beams are used and the range is modified by 2 mm for each beam making the two beam range changes 2 mm and 4 mm as shown in Figure 2. With this technique, OARs are spared unnecessary dose. As an example, we assumed the proton beam to have a uniform RBE of 1.1 and modeled the excess RBE analytically with a hyperbolic tangent centered on the point of maximum dose of the pristine Bragg peak and saturating at 35% with a characteristic length of 2 mm [8]. Applying this model to the individual pristine peaks comprising the delivery of a SOBP allows us to illustrate the changes in plateau flatness, range, and biological effectiveness -- a “range mod” mitigates all three of these effects. With this method, both PSD number and patient set up time potentially decrease. All work presented was conducted in compliance with institutional norms and doses delivered and volumes treated were within the standard of care for the case described.

Figure 1
figure 1

The physical dose for a SOBP composed of four pristine Bragg peaks each separate by 6 mm water equivalent. Applying our illustrative model of increased distal RBE to the individual pristine peaks produces the RBE weighted SOBP. The “range mod” technique mitigates the changes in SOBP plateau flatness, range, and effective dose at the distal edge. Here the modulation is achieved by splitting the SOBP into two parts and shifting one by 3 mm to both smooth out the SOBP and decrease the RBE at the end of the beam.

Figure 2
figure 2

Splitting the SOBP into three beams so as to further reduce the RBE effect. This what is typically done when a single beam plan is being used such as with a posterior fossa boost or germinoma boost after whole ventricular radiation is employed. It can be employed at other times as well when there is significant clinical concern regarding a specific organ at risk.

The “range mod” technique will be illustrated with evaluation of a retreatment patient plan treated at our center with PBT. This patient had recurrent ependymoma in the posterior fossa. Treatment plans were constructed using 3DCRT, IMRT, NRMPBT, and RMPBT techniques. The range modulation plan employed end of range modification of 3 mm, thus avoiding complete transmission through any OAR. While not the case in the given example, our policy as noted above for plans using a single beam is to employ three separate fields, each with a unique range. In the plans shown, two range-modulated fields per beam angle were used.

Xio 6.0 treatment planning software (Elekta AB, Sweden) was used for all cases presented for PBT planning. Active scanning [17, 18] as described previously was employed for the delivery of the spread out Bragg peaks (SOBPs) using apertures and compensators manufactured by IU Health Cyclotron Operations (IUCO). Our uniform active scanning process requires the use of apertures to shape the beam edge and compensators to shape the beam end via direct range compensation, as a sum these pieces of equipment are called patient specific devices (PSDs). Aperture devices were machined out of medical grade brass while compensators were machined from medical grade Lucite using standard procedures; U.S. Food and Drug Administration (FDA) requisite quality assurance was performed. All beam outputs and devices were checked for accuracy before treatment delivery per routine. At our center, each treatment position’s verification images are reviewed by a physician for every fraction in real time either at the gantry or via remote viewing monitor prior to delivering beam.

The DICOM RT data set from the patient’s plan computed on XiO was recovered and imported in the Eclipse 10 (Varian Medical Systems, USA) for side-by-side comparison use. All photon plans were constructed within Eclipse 10. The deployed plan was compared to chart data prior to de-identification in order to confirm the correct recovery of the data and then doses to all contoured structures were converted into percentage format for comparison.

The patient had been previously treated to 54 Gy via coplanar IMRT and relapsed in field. This patient received RMPBT at our center in order to avoid OARs approximately two years ago and is currently doing well without evidence of radiation damage or other local toxicity.

Results and discussion

Multiple beam plans were generated as primary treatment for this patient using NRMPBT (Figure 3), RMBPT (Figure 4), and a comparison of DVH’s for the NRMPBT and RMPBT plans (Figure 5) are shown. Additionally, the actual retreatment plan is shown in order to illustrate how RMPBT is used in this special context and is labeled RMPBTrt (Figure 6). It varies from the optimal RMPBT plan in that a vertex field is used so as to pull dose off the skin and minimize the volume of retreated tissue outside of the PTV. Table 1 outlines the dosimetric comparison of the plans.

Figure 3
figure 3

The plan using NRMPBT.

Figure 4
figure 4

The plan using RMPBT as the primary treatment. DVH colors are the same as used in Figure  3 .

Figure 5
figure 5

Comparison of the DVH’s for several OAR’s between the NRMPBT and the RMPBT plans shown in Figures 3and4respecitively. In every case the RMPBT plan treats less volume of the OAR’s shown.

Figure 6
figure 6

The actual plan delivered using RMPBTrt (as part of retreatment). Vertex beams are used to minimize dose summation with the prior coplanar IMRT plan this patient received. DVH colors are the same as used in Figure 3.

Table 1 Dosimetric comparison of three plans

All planning modalities produced plans that cover the PTV. The RMPBT method treats less total brain than the NRMPBT method given the fact that the beams are not extended to cover the entirety of the brainstem in an effort to avoid ending the beam in the brainstem. When looking at the OAR doses, the difference in plans is pronounced due to this difference. Data exist that suggest doses over 10 Gy are sufficient to ultimately cause hypothalamic dysfunction [19]. In the NRMPBT plan, the average dose is lower, but the peak dose posteriorly is close to the full prescription dose due to the goal of treating through the full brainstem. Only in the RMPBT plan is hypothalamic dose absent completely. This trend continues for the doses to left and right cochleae, the temporal lobes, the pituitary, and the brainstem itself. These data are summarized in Table 1.

As a formal retreatment plan, behind the numbers in the RMPBTrt plan is the concept of treating the previously treated tissue to the lowest sum doses possible. This was achieved by minimizing dose overlap issues between a prior co-planar photon plan and the current retreatment proton plan (shown) via vertex beam usage. The through brainstem approach or NRMPBT plan, had a vertex field been employed, treats a much larger volume of hippocampus and temporal lobe, making its use problematic in the retreatment context.

Even with the RMPBTrt plan’s beam arrangement used to minimize overlap with prior dose, very significant dosimetric saving occurred for the cochleae, the hypothalamus, and the brainstem relative to the NRMPBT plan. The hippocampal dose, despite the vertex field, remained well below the mean dose seen by the other methods.

The value of RMPBT is one of significantly increased patient safety by the direct reduction of treated tissue in a fashion otherwise impossible even for traditional proton therapy because it allows the safe termination of a proton beam or set of beams in an OAR. The technique can be adapted to use in pencil beam planning as well and may prove to be even more critical in that arena as beam edge dosimetry will likely need to become modulated as well.

In each case the number of PSD sets used was decreased or kept the same. We hypothesize that time in the room and complexity was decreased in every range modulated patient scenario relative to electing another angle from which to treat. Range modulation was simple to deliver in the treatment rooms. Finally, as a result of fewer net patient positions being used, fewer verifications films are needed and patient exposure to radiation was decreased.

There are four main advantages to range modulation, or smearing of the distal range of a proton beam, compared to traditional multiple beam proton therapy:

  1. 1.

    Better tissue sparing is achieved via a more aggressive use of distal blocking.

  2. 2.

    Theoretical time-savings in the treatment room as fewer beam angles are needs. This could allow the avoidance of anesthesia in some cases. It also decreases the need to wait for the physicians required to review position films (every field is reviewed every day in our center prior to beam delivery), improving throughput.

  3. 3.

    Less image guidance imaging is used as only the first of a range modified series of beams required image guidance (orthogonal image verification).

  4. 4.

    No new PSD sets have to be manufactured which saves time for the machine shop construction and the cost of the materials and labor involved.

Range modulation as a methodology is not able to solve all problems and can introduce new problems into a plan. There is still entry dose overlap, it adds to the complexity of a plan by adding more beams to a plan, and if too few beams are used skin tolerance can be an issue. Beam angle variation also can be quite valuable to make plans more robust as target and other tissue volumes change during treatment. This is important in anatomical regions containing tissue/air/bone such as the sinuses and hilar regions. This method is a new way of employing the primary principle of radiation safety of “as low as reasonably achievable,” more commonly known by the acronym ALARA, in treatment. It is cost-effective because no new PSD sets need to be constructed and the patient beam angle does not change in the room taking time and requiring set-up imaging for position verification. The presentation of this method is obviously limited by the presentation of only one case, but the idea has translated in our clinic to spinal cord cases, craniopharyngiomas, optic pathway tumors, base of skull tumors, and pelvic tumors. Careful evaluation of the method will demonstrate that the method succeeds by moderately smearing out the sharpness of the end of the SOBP. This modest compromise allows safe stoppage of proton beams within critical structures such as the brainstem as shown. Ultimately it will be up to the treating physician to balance the need for safety against distal blocking goals regarding whether a biologic hot spot in an OAR in a given plan is acceptable.

The method presented in this paper confronts a clinical problem inherent in charged particle therapy – the safe and effective management of the increasing RBE at the end of particle beams. In all forms of past, present, and future of charged particle therapy, the use of this method or an analogous approach will be of use to the clinician when there is an RBE increase at the end of the beam being used. One could even expand this idea to any non-linearity found in RBE in beams as scanned beams allow physicians to compensate for these issues. Our future particle therapy treatment planning will likely also be RBE focused rather than solely energy focused. Table 2 summarizes the method’s usage in our clinic and represents general guidelines.

Table 2 Informal range modulation guidelines employed in our clinic

Conclusions

This present study illustrates a novel method that mitigates the increased RBE at the end of the SOBP in proton treatment planning. It may not be applicable in all situations and decreases the sharpness of the dose fall-off at the end of the SOBP as a result. It has proven practical to design and implement in our clinic. It is most often used in plans using multiple beams. The method represents treatment planning that reflects not only thinking in terms of traditional energy dose (Gy) but also in terms of biologic dose (RBE).

Authors’ information

Each author is associated with the IU Health Proton Therapy Center and the IU School of Medicine. Please see http://iuhealth.org/proton-therapy-center/ and http://radonc.medicine.iu.edu for further information.

Abbreviations

RBE:

Relative biological effectivess

SOBP:

Spread out Bragg peak

PBT:

Proton beam therapy

AU:

Srbitrary units

3DCRT:

Three dimensional conformal radiation therapy

IMRT:

Intensity modulated radiation therapy

NRMPBT:

Non-range modulation proton beam therapy

RMPBT:

Range modulation proton beam therapy

RMPBTrt:

Range modulation proton beam therapy (re-treatment scenario)

OAR:

Organ at risk

Gy:

Gray

PSD:

Patient specific device

FDA:

U.S. Food and Drug Agency

PTV:

Planning target volume

CTV:

Clinical target volume.

References

  1. Hug EB, Muenter MW, Archambeau JO, DeVries A, Liwnicz B, Loredo LN, Grove RI, Slater JD: Conformal proton radiation therapy for pediatric low-grade astrocytomas. Strahlenther Onkol 2002, 178: 10-17. 10.1007/s00066-002-0874-2

    Article  PubMed  Google Scholar 

  2. Chung C, Keating T, Yock T, Tarbell N: Second malignancy risk in patients treated with proton therapy versus conventional photon therapy. In Int J Radiat Oncol Biol Phys 2006, 72: S8.

    Article  Google Scholar 

  3. Zhang R, Howell RM, Giebeler A, Taddei PJ, Mahajan A, Newhauser WD: Comparison of risk of radiogenic second cancer following photon and proton craniospinal irradiation for a pediatric medulloblastoma patient. Phys Med Biol 2013, 58: 807-823. 10.1088/0031-9155/58/4/807

    Article  PubMed Central  CAS  PubMed  Google Scholar 

  4. Taddei PJ, Mirkovic D, Fontenot JD, Giebeler A, Zheng Y, Kornguth D, Mohan R, Newhauser WD: Stray radiation dose and second cancer risk for a pediatric patient receiving craniospinal irradiation with proton beams. Phys Med Biol 2009, 54: 2259-2275. 10.1088/0031-9155/54/8/001

    Article  PubMed Central  PubMed  Google Scholar 

  5. Brodin NP, Munck Af Rosenschold P, Aznar MC, Kiil-Berthelsen A, Vogelius IR, Nilsson P, Lannering B, Bjork-Eriksson T: Radiobiological risk estimates of adverse events and secondary cancer for proton and photon radiation therapy of pediatric medulloblastoma. Acta Oncol 2011, 50: 806-816. 10.3109/0284186X.2011.582514

    Article  PubMed  Google Scholar 

  6. Chung CS, Yock TI, Nelson K, Xu Y, Keating NL, Tarbell NJ: Incidence of second malignancies among patients treated with proton versus photon radiation. Int J Radiat Oncol Biol Phys 2013, 87: 46-52. 10.1016/j.ijrobp.2013.04.030

    Article  CAS  PubMed  Google Scholar 

  7. Paganetti H, Olko P, Kobus H, Becker R, Schmitz T, Waligorski MP, Filges D, Muller-Gartner HW: Calculation of relative biological effectiveness for proton beams using biological weighting functions. Int J Radiat Oncol Biol Phys 1997, 37: 719-729. 10.1016/S0360-3016(96)00540-8

    Article  CAS  PubMed  Google Scholar 

  8. Britten RA, Nazaryan V, Davis LK, Klein SB, Nichiporov D, Mendonca MS, Wolanski M, Nie X, George J, Keppel C: Variations in the RBE for cell killing along the depth-dose profile of a modulated proton therapy beam. Radiat Res 2013, 179: 21-28. 10.1667/RR2737.1

    Article  CAS  PubMed  Google Scholar 

  9. Paganetti H, Niemierko A, Ancukiewicz M, Gerweck LE, Goitein M, Loeffler JS, Suit HD: Relative biological effectiveness (RBE) values for proton beam therapy. Int J Radiat Oncol Biol Phys 2002, 53: 407-421. 10.1016/S0360-3016(02)02754-2

    Article  PubMed  Google Scholar 

  10. Gueulette J, Slabbert JP, Bohm L, De Coster BM, Rosier JF, Octave-Prignot M, Ruifrok A, Schreuder AN, Wambersie A, Scalliet P, Jones DT: Proton RBE for early intestinal tolerance in mice after fractionated irradiation. Radiother Oncol 2001, 61: 177-184. 10.1016/S0167-8140(01)00446-7

    Article  CAS  PubMed  Google Scholar 

  11. Kase Y, Himukai T, Nagano A, Tameshige Y, Minohara S, Matsufuji N, Mizoe J, Fossati P, Hasegawa A, Kanai T: Preliminary calculation of RBE-weighted dose distribution for cerebral radionecrosis in carbon-ion treatment planning. J Radiat Res 2011, 52: 789-796. 10.1269/jrr.11044

    Article  PubMed  Google Scholar 

  12. Blomquist E, Russell KR, Stenerlow B, Montelius A, Grusell E, Carlsson J: Relative biological effectiveness of intermediate energy protons. Comparisons with 60Co gamma-radiation using two cell lines. Radiother Oncol 1993, 28: 44-51. 10.1016/0167-8140(93)90184-A

    Article  CAS  PubMed  Google Scholar 

  13. Paganetti H: Significance and implementation of RBE variations in proton beam therapy. Technol Cancer Res Treat 2003, 2: 413-426.

    Article  CAS  PubMed  Google Scholar 

  14. Jones DTL, Suit HD, Akine Y, Goitein G, Goitein M, Kanematsu N, Maughan RL, Tatsuzaki H, Tsujii H, Vatnitsky SM: Prescribing, recording, and reporting proton-beam therapy. J ICRU 2007,7(No. 2):210. ISBN 1473-6691

    Google Scholar 

  15. Carabe A, Espana S, Grassberger C, Paganetti H: Clinical consequences of relative biological effectiveness variations in proton radiotherapy of the prostate, brain and liver. Phys Med Biol 2013, 58: 2103-2117. 10.1088/0031-9155/58/7/2103

    Article  PubMed  Google Scholar 

  16. Carabe A, Moteabbed M, Depauw N, Schuemann J, Paganetti H: Range uncertainty in proton therapy due to variable biological effectiveness. Phys Med Biol 2012, 57: 1159-1172. 10.1088/0031-9155/57/5/1159

    Article  PubMed  Google Scholar 

  17. Combs SE, Zipp L, Rieken S, Habermehl D, Brons S, Winter M, Haberer T, Debus J, Weber KJ: In vitro evaluation of photon and carbon ion radiotherapy in combination with chemotherapy in glioblastoma cells. Radiat Oncol 2012, 7: 9. 10.1186/1748-717X-7-9

    Article  PubMed Central  PubMed  Google Scholar 

  18. Nichiporov D, Solberg K, Hsi W, Wolanski M, Mascia A, Farr J, Schreuder A: Multichannel detectors for profile measurements in clinical proton fields. Med Phys 2007, 34: 2683-2690. 10.1118/1.2746513

    Article  PubMed  Google Scholar 

  19. Hua C, Wu S, Chemaitilly W, Lukose RC, Merchant TE: Predicting the probability of abnormal stimulated growth hormone response in children after radiotherapy for brain tumors. Int J Radiat Oncol Biol Phys 2012, 84: 990-995. 10.1016/j.ijrobp.2012.01.049

    Article  PubMed Central  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

All authors are funded by Indiana Univeristy and/or the IU Health Proton Therapy Center. No grant funding was used to support this work or for the preparation of this paper.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Jeffrey C Buchsbaum.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

JB invented the idea behind this method, set up its use in the clinic, and was the primary individual behind the interpretation of the plans for patients treated in this fashion. He drafted the paper and served as the main author of the paper. MMcM made substantive intellectual contributions to the idea in this paper and helped draft the paper. PJ made substantive intellectual contributions to the idea in this paper and helped draft the paper. TH constructed the first plans using the method and did the proton therapy plans in this paper. MM contributed to the paper early in the idea formation process via critical discussions with JB. CWC made intellectual contributions to the idea in this paper and helped draft the paper. ID made intellectual contributions to the idea in this paper and helped draft the paper. KM made intellectual contributions to this paper and helped draft the paper. MW made substantive intellectual contributions to the idea in this paper and helped significantly in the drafting of the paper. All authors read and approved the final manuscript.

Authors’ original submitted files for images

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Buchsbaum, J.C., McDonald, M.W., Johnstone, P.A. et al. Range modulation in proton therapy planning: a simple method for mitigating effects of increased relative biological effectiveness at the end-of-range of clinical proton beams. Radiat Oncol 9, 2 (2014). https://doi.org/10.1186/1748-717X-9-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1748-717X-9-2

Keywords