Appropriate patient instructions can reduce prostate motion

Reinhold Graf, Dirk Boehmer, Jacek Nadobny, Volker Budach and Peter Wust*

Author Affiliations

Department of Radiation Oncology, Charité Universitätsmedizin Berlin, Campus Virchow-Klinikum, Augustenburger Platz 1, 13353, Berlin, Germany

For all author emails, please log on.

Radiation Oncology 2012, 7:125 doi:10.1186/1748-717X-7-125

Published: 1 August 2012

Abstract

Background

Interfraction prostate motion must be compensated by increased safety margins. If filling status of rectum and bladder is constant, motion should be reduced. We attempted to reduce interfraction motion errors by proper patient instruction.

Method

In 38 patients pairs of radio-opaque fiducial markers were implanted prior to definitive radiotherapy. Patients were positioned either according to skin marks or infrared body marker. We measured prostate displacement, i.e. pelvic bones versus intraprostatic marker position, via ExacTrac (two orthogonal radiographies) in 1252 fractions. Systematic and random setup and displacement errors were determined and safety margins estimated.

Results

In our study interfraction prostate displacement is < 1 mm in RL direction, and < 2 mm in AP and SI direction. Systematic errors are slightly below random errors (< 1.5 mm). Positioning according skin marks results in higher inaccuracies of ±1.5 – 2 mm in RL and ±2 – 2.5 mm in AP/SI direction.

Conclusions

In case of appropriate patient instructions (constant organ filling) the positioning via bone fusion requires CTV-PTV margins of 2 mm in RL, 4 mm in AP, and 5 mm in SI direction. Studies without any description of patient instruction found much higher margins of > 1 cm in AP and SI direction.